Internal Sensors
Temperature
Type:
DBSA710: PT-100
DBSA720: Digital
Range: -20 to 50 ºC
Resolution: 0.1 ºC
Accuracy:
DBSA710: ±0.3 ˚C
DBSA720: ±0.5 ˚C
Humidity (for DBSA720)
Range: 5 to 95%
Resolution: 0.5%
Accuracy:
± 3% at 10 to 90%
± 5% at 5 to 10%
and 90 to 95%
Supportd Input Types for DBSA710 and DBSA720
4 to 20 mA
Range: 4 to 20 mA
Resolution: 4.76 μA
Accuracy: ±0.5 %
Loop impedance: 21 Ω
Maximum load: 30 mA, 5.2 V
0 to 50 mV
Range: 0 to 50 mV
Resolution: 3 μV
Accuracy: ±0.5 %
Input impedance: 25 MΩ
Maximum voltage: 5.2 V
0 to 1 V
Range: 0 to 1 V
Resolution: 200 μV
Accuracy: ±0.5 %
Input impedance: 25 MΩ
Maximum voltage: 5.2 V
Temperature PT-100 (2-wires)
Range: -200 to 400 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to -60 ˚C ±0.5 %
-60 to 60 ˚C ±0.3 ˚C
60 to 400 ˚C ±0.5 %
Temperature Thermocouple J
Range: -200 to 1,000 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to -60 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
60 to 1,000 ˚C ±0.5 %
Cold junction compensation error: ±0.3 ºC
Temperature Thermocouple K
Range: -200 to 1,000 ºC
Resolution: 0.1 ˚C
Accuracy:
-200 to -60 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
60 to 1,000 ˚C ±0.5 %
Cold junction compensation error: ±0.3 ˚C
Temperature Thermocouple T
Range: -200 to 400 ˚C
Resolution: 0.1 ˚C
Accuracy:
-200 to -60 ˚C ±0.5 %
-60 to 60 ˚C ±0.5 ˚C
60 to 400 ˚C ±0.5 %
Cold junction compensation error: ±0.3 ˚C
Frequency (input 4 only)
Zero crossing detector
Range: 20 Hz to 4 KHz
Input signal: 0 to 4,000 Hz
Input impedance: 470 Ω
Pulse Counter (input 4 only)
Zero crossing detector
Range: 1 to 65,536 counts
Resolution: 1 count
Frequency range: 0 to 4,000 Hz
Input signal: 0 to 5 V
Input impedance: 470 Ω
Contact
Range: Open/Closed
|